1. Applicable room volume: 160m3
2. Environmental temperature: -10°C to 50°C
3. Relative humidity: not exceeding 90%
4. Power source: AC 220V ± 22V
5. Power frequency: 50Hz ± 1Hz
6. Atmospheric pressure: 860-1060hpa;
7. Rated power W: ≤288
8. Airflow rate m³/h: 1600
9. Lifespan of UV lamp: ≥5000 hours
10. Ion concentration: ≥1X106/cm3
11. External dimensions: 520×390×830mm
12. Operating mode: Man-machine interface, cyclic disinfection, program control.
13. The outer surface should be flat, smooth, and free of defects such as scratches, cracks, and rust. The plastic parts should be smooth and free of cracks and deformation. The text labels should be clear and accurate, and the control keys should be sensitive and reliable.
14. The remote control should be able to control the start/stop, time setting, and airflow adjustment.
15. Timing selection: The remote control timer can set the time in 1-24 hour increments, allowing users to choose the disinfection time.
16. The power switch should be clearly recognizable and use text or numbers to indicate whether it is on or off. The timer switch should have a digital time display.
17. Cyclic disinfection airflow: ≥800m3/h
18. The noise level of the disinfectant during operation should be ≤55db(A)
19. In the air disinfection effect test, the killing rate of white staphylococcus aureus (8032 colonies) should be ≥99.90%, or the death rate of natural bacteria should be ≥90% within 2 hours, and it is considered qualified.
20. During disinfection, the ozone concentration in the air should be ≤0.1mg/m³.