1. Applicable room volume: 100m3
2. Ambient temperature: -10℃~50℃.
3. Relative humidity: no more than 90%
4. Power supply: AC 220V±22V
5. Power frequency: 50HZ±1HZ
6. Atmospheric pressure: 860hpa~1060hpa.
7. Rated power W: ≤144
8. Processing air volume m³/h:1000
9. Ultraviolet lamp life: ≥ 5000 hours
10. Negative ion generation: 1 ≥ 1X106 / cm3
11. External dimension: 520×390×830mm
12. Operation mode: man-machine with, cycle disinfection, program control.
13. The outer surface should be smooth and clean, without sharp edges, burrs, cracks and other defects. Plastic spraying parts are bright, uniform color, no shedding, hanging, scratch phenomenon. Plastic parts are bright and clean, no cracking, deformation. Text logo should be clear and accurate, the control keys should be sensitive and reliable.
14. Sterilizer wireless remote control should be able to on/off, time adjustment, wind speed adjustment control, its wireless remote control distance should not be less than 5M.
15. Timing selection: remote control timing key can be set 1-24 hours step amount, according to the user needs to choose the disinfection time.
16. The power switch can clearly identify the on and off, and expressed in words or numbers. Power switch with "1" for on, with "0" for off. Timing switch should have digital time indication.
17. Circulation disinfection air volume: ≥ 800m3 / h
18. Sterilizer working noise ≤ 55db (A)
19. In the air disinfection effect test, the killing rate of white staphylococcus (8032 strains) ≥ 99.90%, or ≥ 90% of the natural bacterial extinction rate of the disinfection time of no more than 2h is qualified.
20. The amount of ozone in the air during disinfection should be ≤0.1mg/m³.